Main Function
Zatha series use a principle that works cutting and coagulation of cellular tissue via high-frequency current through the body contact load and resistance by using the heat generated.
As the high-frequency current through the body, heat will be occurred by heat and arc discharge in active electrode and return-pad fitted wide area in the body. So, the heat makes heating the cellular tissue and it is possible to cut and coagulate as operation.
Return-pad and the active electrode are mono-polar in combination. In another function, the end of electrode as shape of forceps is a pair of electrode. The high frequency current works among the electrode in the forceps. It called a bipolar.
Feature
Safety
In the case, the connection is incorrect between patients and attached return-pad with a patient as operation, automatically RF output will be blocked and sound an alarm.
Preciseness
All the operations of ZATHA Series are digitalized output 1W of unit by means of a microprocessor. So, more precise operation is possible. Also user turns off a unit, the unit memories final setting value in last operation.
Specification
ZATHA-S | |
Electrical specifications | |
Current source | 100 ~ 230V |
Repeat | 50/60 Hz |
The power input is over the maximum. | 1100v Ampere |
Fuse | T10A |
Power output specification | |
Maximum cutting limit | 300 watts |
Maximum mix | 200 watts |
Maximum amount of COAG | 200 watts |
Maximum bipolar | 100 watts |
Dimension and weight | |
Width × depth × height | 330x350x130 ملم |
Net weight | 7.5 kg |