Feature:
- Operate the evaporator only with dry medical gases
- MSS 3 looks like Tec 3 but here the similarity ends.
- Over the past ten years, we have made many improvements to the original design.
- In addition to those described below (Tec 3).
- We have created a more stable evaporator that gives the output accuracy between +/- 2% and 5% of the connection setting, as well as a longer period of “duration of use”.
Features
- Titration of the selected anesthetic drug.
- 0.25 to 5% (Sevoflurane 8%) was calibrated using a “laser refractometer” (Tec 3 0.5 to 5.0% No 8% Sevoflurane).
- Temperature, flow and duration of use.
- A security lock to help prevent accidental operation of the disc. Customize
- Calibrate at 22 degrees Celsius and 35 degrees Celsius + / - 1 degree Celsius.
- Calibration range: 0.25% - 0.5% over 5% (Sevoflurane 8%) V.V. (Tec 3 0.5 إلى 5٪ No 8٪ Sevoflurane)).
- Fluctuations in temperature, flow rate and duration of use are automatically compensated during use.
- When used with I.P.P.V, the difference in output is minimal.
- Nominal working range: fresh gas flow – 200 cc to 15 l / min.
- Ambient temperature – 18 degrees Celsius to 35 degrees Celsius.
Gas flow resistance:
- 5 cm Weight. In the “OFF” mode at 5 l / min oxygen.
- 21-29 CM Wg. When pumping steam at 5 l / min oxygen is pumped at 21 degrees Celsius.
Liquid capacity:
- The amount of anesthetic drug for a full charge of the vaporizer is 150 ml (Tec 3 125ml).
- The amount that the wick system retains-35 ml.